Your browser doesn't support javascript.
loading
A Novel Simultaneous Determination of Sarpogrelate and its Active Metabolite (M-1) in Human Plasma, Using Liquid Chromatography-Tandem Mass Spectrometry: Clinical Application
Annals of Laboratory Medicine ; : 391-398, 2015.
Article in English | WPRIM | ID: wpr-57046
ABSTRACT

BACKGROUND:

This study describes a novel analytical method for simultaneously determining sarpogrelate and its metabolite (M-1) in human plasma, using liquid chromatography coupled with tandem mass spectrometry, with electrospray ionization in the positive ion mode.

METHODS:

Sarpogrelate, M-1, and labeled internal standard (d3-sarpogrelate) were extracted from 50 microL of human plasma by simple protein precipitation. Chromatographic separation was performed by using a linear gradient elution of a mobile phase involving water-formic acid (99.90.1, v/v) and acetonitrile-formic acid (99.90.1, v/v) over 4 min of run time on a column, with a core-shell-type stationary phase (Kinetex C18, 50 mm x 2.1 mm i.d., 2.6-microm particle size, Phenomenex, USA). Detection of the column effluent was performed by using a triple-quadruple mass spectrometer in the multiple-reaction monitoring mode.

RESULTS:

The developed method was validated in human plasma, with lower limits of quantification of 10 ng/mL for sarpogrelate and 2 ng/mL for M-1. The calibration curves of sarpogrelate and M-1 were linear over the concentration ranges of 10-2,000 and 2-400 ng/mL, respectively (R2>0.99). The carry-over effect, precision, accuracy, and stability of the method met the criteria for acceptance.

CONCLUSIONS:

A simple, fast, robust, and reliable analytical method was successfully developed and applied to the high-throughput determination of sarpogrelate and its metabolite in real plasma samples in a pharmacokinetic study of healthy subjects.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Particle Size / Plasma / Mass Spectrometry / Calibration / Chromatography, Liquid / Tandem Mass Spectrometry Limits: Humans Language: English Journal: Annals of Laboratory Medicine Year: 2015 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Particle Size / Plasma / Mass Spectrometry / Calibration / Chromatography, Liquid / Tandem Mass Spectrometry Limits: Humans Language: English Journal: Annals of Laboratory Medicine Year: 2015 Type: Article