Your browser doesn't support javascript.
loading
Improved RP-HPLC method for determination of ferulic acid concentration in human serum / 中草药
Chinese Traditional and Herbal Drugs ; (24)1994.
Article in Chinese | WPRIM | ID: wpr-570665
ABSTRACT
Object To develop an improved RP-HPLC method for determining ferulic acid in the human serum. Methods The determination was carried on RP-HPLC, using Kromasil-C 18column, methanol-water-acetic acid (36.4∶63∶0.6) as mobile phase with a flow rate of 1.0 mL/min and at the detection wavelength of 322 nm, and p-hydroxybenzaldehyde was used as internal standard. The serum sample was deproteinized with acetonitrile to extract ferulic acid. Results The calibration curve showed good linearity over the range of 9.94-159.04 ng/mL (r=0.992 5). RSD was less than 10% within day and day-to-day, the average recovery was 99.77% and the minimal concentration in serum was 5 ng/mL. Conclusion This method, which is simple, rapid, sensitive, reproducible and low toxic, is appropriate for the quantitative determination of ferulic acid in the human serum.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Traditional and Herbal Drugs Year: 1994 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Traditional and Herbal Drugs Year: 1994 Type: Article