Effect of Storing Time of Venous Blood Samples on the Differential Counts of White Blood Cells by Sysmex XE-2100 Hematology Analyzer / 热带医学杂志
Journal of Tropical Medicine ; (12): 786-789,799, 2005.
Article in Chinese | WPRIM | ID: wpr-571786
ABSTRACT
Objective To evaluate the influence of the storing time of venous blood samples on the differential count of WBC by Sysmex XE-2100 hematology analyzer. Methods At room temperature, the precision of the differential count of WBC by Sysmex XE-2100 hematology analyzer were tested. 38 samples were taken the differential count of WBC by Sysmex XE-2100 hematology analyzer after stored for 0, 2, 4, 8, 24 and 48 h. Differential count of WBC was also taken under microscope for comparison. Results The precision of differential count of WBC by Sysmex XE-2100for all the samples was in the allowable range. The correlation coefficient of the differential count of WBC by two methods for neutrophils, lymphocytes, monocytes, eosinophils and basophils were 0.9859, 0.9775, 0.8053, 0.8695and 0.5243 (P<0.01). There was insignificant difference in the test at 8h, very significant difference of MONO and EOS at 48 h. Differences of EOS, MONO between two methods were significant increased at 8 h. Conclusion At room temperature, the differential count of WBC of venous blood samples by Sysmex XE-2100 hematology analyzer should finish within 8 h.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Journal of Tropical Medicine Year: 2005 Type: Article