Studies on standardization of serum enzyme determination / 中华检验医学杂志
Chinese Journal of Laboratory Medicine
;
(12)2001.
Article
in Chinese
| WPRIM
| ID: wpr-582467
ABSTRACT
Objective Investigating for the standardization of serum enzyme determination by using enzyme calibrator in lab tests Methods 150 laboratories were involved in this program. One enzyme calibrator and three patient samples were send to each participate lab The calibrator was tested using routine method and record the results The patient samples were analyzed before and after calibration Record the results All lab data was processed with computer Results Compare with the target values, the bias of alanine aminotransferase (ALT), aspartate aminotransferase (AST), amylase (AMY) and lactate dehydrogenase (LD) assays were 3 8%, -1 8%, 2 3% and -5 2% respectively, the bias of alkaline phosphatase (ALP), gamma glutamyltransferase (GGT) and creatine kinase (CK) assays were 13 7%, -13 9 and -19 2% respectively The precision of the enzyme assays were improved by using calibrators Conclusions The traceability of ALT, AST, AMY and LD assays were reasonable Calibrating by using enzyme calibrator can improve the assay precision among laboratories
Full text:
Available
Index:
WPRIM (Western Pacific)
Language:
Chinese
Journal:
Chinese Journal of Laboratory Medicine
Year:
2001
Type:
Article
Similar
MEDLINE
...
LILACS
LIS