Your browser doesn't support javascript.
loading
Research of effect of stray capacitance on measurement accuracy in tissue impedance spectroscopy measurement / 医疗卫生装备
Chinese Medical Equipment Journal ; (6)1993.
Article in Chinese | WPRIM | ID: wpr-583779
ABSTRACT
In the tissue impedance spectroscopy measurements ,four-electrode method are normally used. A measurement error caused by stray capacitance between lines of drive electrode and receive electrode is not a trivial problem. This paper studied this error and reached the results This stray capacitance introduces errors which increase with frequency and has more effect on the imaginary part than on the real part in the measurement frequency range.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Medical Equipment Journal Year: 1993 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Medical Equipment Journal Year: 1993 Type: Article