Your browser doesn't support javascript.
loading
Transfering the target values from the WHO-IFCC reference material for apolipoproteins A1 and B to in-house calibrator / 中华检验医学杂志
Chinese Journal of Laboratory Medicine ; (12)2001.
Article in Chinese | WPRIM | ID: wpr-584696
ABSTRACT
Objective To prepare in-house calibrator of apolipoprotein A1 and apolipoprotein B with the values are traceable to the WHO-International reference materials SP1-01 and SP3-07.Methods We participated in the Northwest lipid research laboratory (NWLRL) protocol for target value transfer of WHO-IFCC apo A1 and apoB reference materials, and evaluated of comparability of the measurement.The analyses were performed on a Olympus AU 400, apo A1 and apoB were determined by immunoturbidimetric endpoint method.Results Through 3-step study, the precision and accuracy of apo A1 and apoB determination of our company met the NWLRL criteria.Correlation coefficients between obtained and assigned values of NWLRL on 40 individual serum samples were 0.983 for apo A1, 0.987 for apoB, the average absolute bias were 2.7 % for apo A1, 3.0 % for apoB, respectively.Conclusion NWLRL issued the certificate to Zhongsheng Beikong Bio-Technology and Science Inc.for the in-house apo A1 and apoB calibrator values are traceable to the WHO- International reference materials.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Laboratory Medicine Year: 2001 Type: Article