ABSTRACT
The efficiency of Siemens E.CAM is evaluated by analyzing and practicing the technology of Siemens U-fly.Basic testing procedure is run.Standard source 99mTc and 57Co is used respectively and the operation is performed according to the instruction.The detecting head's uniformity can be corrected in this way with satisfactory results.It is suggested uniformity should be tested every day before study because it is the most important factor for the quality of clinical images.