Your browser doesn't support javascript.
loading
Decrease in Pulse Oximeter Readings Following Injection of Isosulfan Blue Dye / 한국유방암학회지
Journal of Korean Breast Cancer Society ; : 24-28, 2003.
Article in Korean | WPRIM | ID: wpr-58801
ABSTRACT

PURPOSE:

A sentinel lymph node mapping with blue dye has been well accepted as a common procedure in breast cancer surgery. However, it is well known that blue dye absorbed into the circulation may interfere pulse oximetery reading. The aim of this study was to evaluate the change of pulse oximetery reading by isosulfan blue dye injection during sentinel lymph node mapping.

METHODS:

Thirteen breast cancer patients with normal preoperative cardiopulmonary functions were studied. Four ml of isosulfan blue dye was injected subdermally when the patient became stable after induction of general anesthesia. The pulse oximetery was monitored continuously. Multiple arterial blood gas analyses (ABGA) were performed before dye injection and 10, 30, 40 minutes after dye injection. The results of oxygen saturturation by oximetery (SpO2) and the results of arterial oxygen tension (SaO2) and arterial oxygen saturation (SaO2) by ABGA were compared.

RESULTS:

The value of both SaO2 and PaO2 measured by ABGA has not been altered by isosulfan dye injection. However SpO2 decreased by isosulfan dye injection. SpO2 decrease started 8.2+/-1.5 (2~0) minutes after dye injection and returned to preinjection level by 85.7+/-5.6 (60~126) minutes after injection. The lowest vaule of SpO2 was 95.6+/-1.2% (93~97). Mean duration of SpO2 decrease was 77.5+/-6.2 (40~117) minutes. The duration of SpO2 decrease was longer in the aged patients, but it was not statistically significant (p=0.3). There was no siginificant difference in duration of SpO2 decrease according to injection site, operation method, and body mass index (BMI).

CONCLUSION:

.Isosulfan dye injection using for sentinel lymph node mapping causes no change in true ABGA results but causes a mild reversible decrease in SpO2, It is important to look for other causes when SpO2 decrease is significant and persistent.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Oxygen / Reading / Blood Gas Analysis / Breast Neoplasms / Body Mass Index / Anesthesia, General / Lymph Nodes Limits: Humans Language: Korean Journal: Journal of Korean Breast Cancer Society Year: 2003 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Oxygen / Reading / Blood Gas Analysis / Breast Neoplasms / Body Mass Index / Anesthesia, General / Lymph Nodes Limits: Humans Language: Korean Journal: Journal of Korean Breast Cancer Society Year: 2003 Type: Article