Your browser doesn't support javascript.
loading
Certification of linear range of WBC measured by CD3700 and application in ISO15189 / 基础医学与临床
Basic & Clinical Medicine ; (12)2006.
Article in Chinese | WPRIM | ID: wpr-590013
ABSTRACT
Objective To certify the linear range of WBC measured by CD3700.Methods Use the white stratum of blood after centrifugation.Then,the high value specimens were diluted with saline to get a series of samples with different concentrations.Every specimen was measured twice in order to keep the sequence of specimens randomly.The perform outlier test,polynomial regression analysis,and the random error estimation were carried out.Results There was no outlier in the results.It was demonstrated by polynomial regression to be linear from(0~219)?109 L-1.The random error was within the allowable limit of 1/4CLIA'88(3.75%).Conclusion The linear range of WBC measured by CD3700 is just like that of the manufacturer claims.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Basic & Clinical Medicine Year: 2006 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Basic & Clinical Medicine Year: 2006 Type: Article