Performance Tests and Quality Control of Siemens ECAT EXACT HR+ PET / 医疗卫生装备
Chinese Medical Equipment Journal ; (6)2004.
Article in Chinese | WPRIM | ID: wpr-592091
ABSTRACT
Objective To evaluate the performance parameters of the PET instrument, which has been used for years in our hospital, assuring the stability of image quality and performance parameters. Methods NEMA NU 2-1994 Standards was used as reference for performance measurement of PET, which were published by National Electrical Manufacturers Association(NEMA). The NEMA standard cylinder model, 18F radiation source, and 18F and 68Ge columniform radiation source were used. Results Projects for performance measurement and daily quality control were drawn up. Conclusion This project provides a reference and methods for the performance measurement of PET instrument and daily quality control. It better ensures the stability of image quality and performance parameters of the PET instrument.