Your browser doesn't support javascript.
loading
Metrology Measurement and Quality Analysis of Medical Oxygen Flowmeter / 医疗卫生装备
Chinese Medical Equipment Journal ; (6)2003.
Article in Chinese | WPRIM | ID: wpr-592781
ABSTRACT
Objective To check some medical oxygen flowmeters in use in the military area command. Methods According to JJG 917-1996 buoy oxygen inhalant measuring rules, 930 oxygen flowmeters in the military area command were measured. Results The unqualified reasons of medical oxygen flowmeter mainly lay in the capability decline of pressure resistance of humid bottle, the poor airproof of the interface, the overflow of pressure, and the overflow of flux value. Conclusion To improve the quality situation of medical oxygen flowmeter, the department of technical supervision, hospital and manufacturer need to make efforts together.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Medical Equipment Journal Year: 2003 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Medical Equipment Journal Year: 2003 Type: Article