A Respiratory Signal Detection Circuit with Automatic Compensation Function / 医疗卫生装备
Chinese Medical Equipment Journal
;
(6)1989.
Article
in Chinese
| WPRIM
| ID: wpr-594174
ABSTRACT
Objective To introduce a new kind of circuit with temperature and voltage double compensation function in view of the deficiency of ambient temperature change on existing respiratory signal detection circuit. Methods The respiratory signal is detected by the characteristics of diode junction voltage with temperature changes. Results The functions of automatic temperatures compensation and automatic voltage fluctuation compensation are realized with the differential connection of two identical diodes. Conclusion The experimental results shows that the proposed circuit is basically immune to the fluctuation of ambient temperature changes and voltage fluctuation, its structure is simple and the accuracy of breathing signal detection is also improved.