Your browser doesn't support javascript.
loading
Performance Evaluation of BC-5500 Hematology Analyzer / 医疗卫生装备
Article in Zh | WPRIM | ID: wpr-594997
Responsible library: WPRO
ABSTRACT
0.99). DC: reproducibility was good for neutrophils, lymphocytes monocytes, eosinophils and basophils. Comparison of the results by instrument with manual for normal samples in morphology, the correlation was better for neutrophils, lymphocytes and eosinophils (r:0.968~0.983) ,good for monocytes(r=0.917), not good for basophils(r=0.659);The WBC scattergram would change and alarm flags would display when there are neutrophilic stab granulocytes, abnormal or atypical lymphocytes and immature cells in higher proportion.Conclusions The analyzer can be used to test blood cell parameters accurately and reliably. Its main performance indices accorded with the experimental requirements; The results were credible. It is necessary to check with microscopy for DC before reported when it were doubtted.
Key words
Full text: 1 Index: WPRIM Language: Zh Journal: Chinese Medical Equipment Journal Year: 1993 Type: Article
Full text: 1 Index: WPRIM Language: Zh Journal: Chinese Medical Equipment Journal Year: 1993 Type: Article