New-type Dynamic Oxygen Inhalation Display Method for Hyperbaric Oxygen Chamber / 医疗卫生装备
Chinese Medical Equipment Journal
;
(6)2003.
Article
in Chinese
| WPRIM
| ID: wpr-596604
ABSTRACT
The electronic gas flow meter is used to realize dynamic oxygen inhalation display for hyperbaric oxygen chamber,then such problems can be conquered as high-resistance oxygen inhalation with glass rotameter and risks for adopting second-class oxygen regulator with sensors.With a low oxygen inhalation resistance and detection of oxygen inhalation state outside the chamber,the accidents can be avoided when hyperbaric oxygen chamber applied.