Your browser doesn't support javascript.
loading
An efficiency calibration method without a radioactive source using BOMAB phantom and Monte Carlo simulation for Inspector 2000 gamma spectroscopy system / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 432-434,448, 2009.
Article in Chinese | WPRIM | ID: wpr-597290
ABSTRACT
Objective To establish an efficiency calibration method without a radioactive source for Inspector 2000 gamma spectroscopy system.Methods At the fwst step,geometry parameters of the detector were properly specified by comparisons of FEPE(Full Energy Peak Efficiency)between Monte Carlo(MC)calculations and the average measured values of HPGe and NaI detectors using point sources(137 Cs).The differences between calculations and measurements were generally within ±10% for gamma rays.Then,Monte Carlo simulation was used to evaluate the counting efficiency of Nal(TI)and HPGe detectors for BOMAB phantoms.Corresponding efficiency curves and functions were fitted.Results The counting efficiency received from the fitted efficiency functions had a good agreement with those from MC simulation.The bias for Nal detector ranged from-19% to 18%,while the bias for HPGe detector ranged from-11% to 17%.The above errors were totally acceptable in the on-site deployment during nuclear and radiological emergency events.Conclusions Monte Carlo method is simple,time and laborsaving in determing the counting efficiency of gamma spectroscopy system.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Prognostic study Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2009 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Prognostic study Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2009 Type: Article