Your browser doesn't support javascript.
loading
The Evaluation of HRTII in Screening of Glaucoma with Different Discrimiant Formulas in Koreans
Journal of the Korean Ophthalmological Society ; : 94-98, 2004.
Article in Korean | WPRIM | ID: wpr-59764
ABSTRACT

PURPOSE:

To find the optimal discrimiant formula of HRTII program in detection of glaucoma in Korean patients.

METHODS:

The objects of this study were 421 eyes of 225 patients at Kangnam St. Mary's Hospital from December in 2002 to February in 2003. Humphery automated perimetry, program C24-2 testing, and HRTII (Software IP1-V1.6) were performed. The results were analyzed with Moorfields discrimiant formula, Mikelberg discrimiant formula and Park discrimiant formula.

RESULTS:

In 421 eyes, the number of glaucomatous eyes was 214. The accuracy of Moorfields discrimiant formula was 71.25%, its sensitivity was 72.42%, and its specificity was 70.04%, The accuracy of Mikelberg discrimiant formula was 71.49%, its sensitivity was 71.96%, and its specificity was 71.01%, The accuracy of Park discrimiant formula was 72.92%, its sensitivity was 73.31%, and its specificity was 72.46%.

CONCLUSIONS:

Park discrimiant formula was expected to be useful in the screenning of glaucoma in Korean patients.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Glaucoma / Mass Screening / Sensitivity and Specificity / Visual Field Tests Type of study: Diagnostic study / Screening study Limits: Humans Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 2004 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Glaucoma / Mass Screening / Sensitivity and Specificity / Visual Field Tests Type of study: Diagnostic study / Screening study Limits: Humans Language: Korean Journal: Journal of the Korean Ophthalmological Society Year: 2004 Type: Article