Your browser doesn't support javascript.
loading
Research on the effect of contrast Shimadzu CL8000 Hitachi 7180 automatic biochemical analyzer cleaning / 中国医学装备
China Medical Equipment ; (12): 36-37,38, 2014.
Article in Chinese | WPRIM | ID: wpr-598801
ABSTRACT

Objective:

To me affiliated hospitals are currently using Shimadzu CL8000 Hitachi 7180 automatic biochemical analyzer comparison study on Shimadzu CL8000 Hitachi 7180 automatic biochemical analyzer cleaning effect for clinical obtain accurate test results provide a reliable basis.

Methods:

Patients with the same blood samples TG, TBA testing, using different instruments, different sequence of tests in order to determine Hitachi 7180 automatic biochemical analyzer and Shimadzu CL8000 automatic biochemical analyzer test results whether they carry the previous reagents, according to the detection results to determine two kinds of automatic biochemical analyzer after cleaning effect.

Results:

A group of TBA measurements were significantly higher than in group B, C group, D group(x2=4.091, P<0.05), comparing the results with statistical significance;B group, C group, D group compared TBA measurements no significant difference, the difference was not statistically significant.

Conclusion:

Shimadzu CL8000 automatic biochemical analyzer cleaning better than Hitachi 7180 automatic biochemical analyzer, its stable performance, excellent cleaning effect, clinicians can provide accurate diagnosis and treatment basis, improve patient outcomes, quality of life and protect their life safety, may be of clinical use.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Medical Equipment Year: 2014 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Medical Equipment Year: 2014 Type: Article