Your browser doesn't support javascript.
loading
Explore the feasibility of preventive maintenance model of hemodialysis machine based on quality control detection / 中国医学装备
China Medical Equipment ; (12): 60-61,62, 2014.
Article in Chinese | WPRIM | ID: wpr-600201
ABSTRACT

Objective:

To investigate the feasibility of hemodialysis machine preventive maintenance model of based on quality control detection.

Methods:

Selected sample of 32 hemodialysis machines of two major imported brands, 16 of them using the traditional preventive maintenance mode for maintenance, and the other 16 sets of quality control detection based preventive maintenance hemodialysis machine maintenance mode. That by use of HDM99XP detector instrument for testing the five hemodialysis machine main indicators, maintenance cycle is 16 weeks, and then the equipment failure rate indicators of two modes of preventive maintenance were compared.

Results:

The hemodialysis machine preventive maintenance model of based on quality control detection rise superior to the traditional maintenance mode in lower failure rate, quantitative, objective, and less affected by human factors.

Conclusion:

The hemodialysis machine preventive maintenance model of based on quality control detection was better than the traditional preventive maintenance mode,also a promising preventive maintenance mode.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study / Prognostic study Language: Chinese Journal: China Medical Equipment Year: 2014 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study / Prognostic study Language: Chinese Journal: China Medical Equipment Year: 2014 Type: Article