A comparative analysis on results of blood glucose detected by POCT portable glucose meters and automatic biochemical analyzer / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 734-735,737, 2015.
Article in Chinese | WPRIM | ID: wpr-600566
ABSTRACT
Objective To compare the results of blood glucose detected by the POCT glucose meters and the automatic biochem‐ical analyzer ,and to periodically evaluate the reliability of the portable glucose meters .Methods 22 patients with fasting blood‐glu‐cose of 0 .88 - 30 .82 mmol/L were selected and the NaF anticoagulation whole blood was collected .Each sample were detected for 3 times .The mean value was taken for conducting the bias and correlation evaluation .Results The bias of 20 glucose meters was within ± 0 .2 and PT > 80% ,and the bias of 5 glucose meters ± 0 .2 in 2 - 6 concentration values .Conclusion The testing results of 20 glucose meters are identical with those detected by the biochemical analyzer and meet the accuracy standards stipulated by the Ministry of Public Health ,and 5 glucose meters have relatively big differences .