Your browser doesn't support javascript.
loading
Comparison analysis between different analysis systems for the detection of electrolytes / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 1108-1109,1112, 2015.
Article in Chinese | WPRIM | ID: wpr-601061
ABSTRACT
Objective To investigate the comparability of K + ,Na+ and Cl- measurement between Cobas8000 analyzer and AU640 analyzer .Methods According to the EP9‐A2 document of the National Committee for Clinical Laboratory Standards ,the precision of AU640 analyzer for the detection of K + ,Na+ and Cl- was evaluated ,and the correlation between the detected results of the two analyzers were also analyzed .Results AU640 analyzer were with high precision for the detection of K + ,Na+ and Cl- ,in‐tra‐assay and inter‐assay coefficient of variation of which were both less than 2% .The correlation coefficients of the two analyzers were higher than 0 .975 .Conclusion Cobas8000 analyzer and AU640 analyzer might be with fine comparability for the detection of K+ 、Na+ and Cl- ,both of which could provide accurate detected results .

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study / Practice guideline Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2015 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study / Practice guideline Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2015 Type: Article