ABSTRACT

Objective:

To establish a method for the determination of trace nickel in agomelatine.

Methods:

An inductive coupled plasma atomic emission spectrometry method was applied in the determination at 221. 648 nm. The sample solution was prepared by the ignited residue of agomelatine. The content of nickel was determined using the standard curve.

Results:

The linear range was 0. 025-1. 000 μg·ml-1(r=0. 999 8). The RSD of precision was 0. 63%. The detection limit was 0. 000 8μg·ml-1. The quantitative limit was 0. 003μg·ml-1 . The average recovery was 99. 4% with RSD of 2. 20%. The RSD of repeatability was 1. 33%.

Conclusion:

The method is simple, sensitive and accurate in the determination of trace nickel in agomelatine.