Your browser doesn't support javascript.
loading
Establishment and validation of finger-ring dosimeter standard calibration quality / 中国医学装备
China Medical Equipment ; (12): 10-12,13, 2016.
Article in Chinese | WPRIM | ID: wpr-603931
ABSTRACT

Objective:

According to the technical standard of ISO and current regulations in China, the X/γ standard calibration quality for the finger-ring dosimeter irradiation was established in Beijing SSDL, it can ensure the traceability of surveillance data within countrywide.

Methods:

Air kerma-rate was determined in the X-ray beams,137Cs and60Co by a standard dosimeter, combined transfer coefficientHp(0.07) provided by standards, and confirmed condition of X/γ standard calibration quality for the finger-ring dosimeter irradiation. Using finger-ring TLD dosimeter verified the radiation beam qualities Hp on the ISO finger-phantom.

Results:

The linearity and energy response of the TLD dosimeter met ISO and national standard in Beijing SSDL.

Conclusion:

The X/γ standard radiation field can be used for the calibration of finger-ring dosimeter standard curves and experiment of research, and can further promote the comparability and reliability of the surveillance data.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Prognostic study Language: Chinese Journal: China Medical Equipment Year: 2016 Type: Article