Your browser doesn't support javascript.
loading
The simulation and optimization of front window airflow in biological safety cabinets / 中国医学装备
China Medical Equipment ; (12): 17-18,19, 2016.
Article in Chinese | WPRIM | ID: wpr-604280
ABSTRACT

Objective:

In order to explore the maximum and the optimal velocity of front window air in biological safety cabinets and provide a scientific evidence for relevant test.

Methods:

As classⅡ biological safety cabinet an testing sample, computational fluid dynamics method is adopted to establish numerical model, then made the simulation of front window airflow.

Results:

The velocity relationship between front window airflow and the internal air has been confirmed.

Conclusion:

The maximum velocity of the front window flow is 1.4 m/s and the optimal velocity is 1.0 m/s. The tested data and conclusion can give some suggestions and reference for biological safety cabinets detection.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Medical Equipment Year: 2016 Type: Article

Similar

MEDLINE

...
LILACS

LIS