The simulation and optimization of front window airflow in biological safety cabinets / 中国医学装备
China Medical Equipment
;
(12): 17-18,19, 2016.
Article
in Chinese
| WPRIM
| ID: wpr-604280
ABSTRACT
Objective:
In order to explore the maximum and the optimal velocity of front window air in biological safety cabinets and provide a scientific evidence for relevant test.Methods:
As classⅡ biological safety cabinet an testing sample, computational fluid dynamics method is adopted to establish numerical model, then made the simulation of front window airflow.Results:
The velocity relationship between front window airflow and the internal air has been confirmed.Conclusion:
The maximum velocity of the front window flow is 1.4 m/s and the optimal velocity is 1.0 m/s. The tested data and conclusion can give some suggestions and reference for biological safety cabinets detection.