Your browser doesn't support javascript.
loading
Evaluation on measurement uncertainty of detecting procedure of amino acids and carnitines by using non-derivatized tandem mass spectrometry method / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 2423-2425,2428, 2016.
Article in Chinese | WPRIM | ID: wpr-604620
ABSTRACT
Objective To evaluate the measurement uncertainty of the detecting procedure of amino acids and carnitines by the Waters ACQUITY TQD tandem mass spectrometer and PerkinElmer NeoBaseTM non‐derivatized MSMS kit ,and to discuss the meaning of evaluation .Methods According to the method provided by the Medical Laboratories‐Evaluation and Expression of Measurement Uncertainty ,the internal quality control was detected by using two different batch codes of kit for 20 d ,once before and after the routine sample detection on each working day ,the detector were randomly changed ,and then the measurement uncer‐tainty introduced by measurement reproducibility was calculated;the amino acid external assessment quality control data in 20 times of neonatal hereditary metabolic disease tandem mass spectrometry screening provided by the National Center for Clinical Laborato‐ry in 2014 and 2015 ,and the same and 16 times of carnitines external assessment quality control data were performed the statistics , and then the measurement uncertainty introduced by bias was calculated .Next the relative standard uncertainty and the relative ex‐panded uncertainty according to the measurement uncertainty introduced by bias and measurement reproducibility were calculated . The allowed total error indicators of amino acid and carnitines external quality assessment in the neonatal hereditary metabolic dis‐ease tandem mass spectrometry screening by the National Center for Clinical Laboratory were used as the target expanded uncer‐tainty .Results The relative expanded uncertainties of citrulline ,methionine ,phenylalanine ,propionyl carnitine ,octanoyl carnitine , dodecanoyl carnitine ,palmitoyl carnitine and octadecanoyl carnitine were 19 .1% -26 .1% (k=2) ,which were smaller than the tar‐get uncertainty .The relative expanded uncertainties of leucine ,tyrosine ,valine ,free carnitine were 31 .0% -43 .3% (k=2) ,which were greater than the target uncertainty .The uncertainty of isovaleryl carnitine needed to be estimated separately .Conclusion As‐sessing the measurement uncertainty of the detecting procedure of amino acids and carnitines by the non‐derivatized tandem mass spectrometer method can not only provide an opportunity for continuously improving the detection quality ,but also can help the ex‐perimental technique staffs to interpret the test data correctly and the clinician to use the detection reports correctly .

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2016 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2016 Type: Article