Your browser doesn't support javascript.
loading
Technique in linear accelerator total body irradiation / 中国医学物理学杂志
Chinese Journal of Medical Physics ; (6): 18-19,26, 2001.
Article in Chinese | WPRIM | ID: wpr-605008
ABSTRACT
This article describes the physical, technical, and dosimetric aspects of total body irradiation (TBI) that was carried out by using 6MV X-Ray from Varian 2300 C/D Linear Accelerator at a distance of 450 cm from target to the treatment table and at a gantry angle of 270°.The dose to lung tissue was limit by setting the individual lead compensators customized before, and using DPD-510 to monitor the absorbed dose of the reference point the absorbed dose in depth of half of body will be (Din+Dout)/2 after taking treatment in both AP position and PA position.
Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Medical Physics Year: 2001 Type: Article