Your browser doesn't support javascript.
loading
Immunosensor construction of label free detection of CD4+T lymphocytes / 国际检验医学杂志
International Journal of Laboratory Medicine ; (12): 753-755,759, 2017.
Article in Chinese | WPRIM | ID: wpr-606794
ABSTRACT
Objective To construct an immunosensor for detecting CD4+ T lymphocytes without labeling .Methods The staphy-lococcus protein A(SPA) method was adopted to conduct the oriented immobilization of CD4 monoclonal antibodies on the gold in-terdigitated microelectrode surface for capturing CD4+ T lymphocytes .Then cyclic voltammetry(CV) method was used to conduct the representation of modification situation on the gold interdigitated microelectrode surface .Finally the electrochemical impedance spectroscopy(EIS) was used to detect the impedance of CD4+ T lymphocytes captured by the immunosensor .The standard curve was drawn by the impedance values change obtained by the equivalent electric circuit fitting .Results The linear range of this im-munosensor for detecting CD4+ T lymphocytes was (5 × 103 -5 .0 × 106 )/mL ,with lower detection limit of 5 .0 × 102/mL .Conclu-sion The constructed immunosensor has accurate and reliable detection results uhidn is simple to operate accurate ,convenient and cheap ,which might be expected to be used in the real-time detection system ,and offers help for realizing rapid ,accurate and inex-pensive CD4+ T lymphocyte count .

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2017 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study Language: Chinese Journal: International Journal of Laboratory Medicine Year: 2017 Type: Article