Your browser doesn't support javascript.
loading
Study on step accuracy measurement and dose deviation of 252 Cf neutron afterloading radiotherapy machine / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 626-629, 2017.
Article in Chinese | WPRIM | ID: wpr-611154
ABSTRACT
Objective To study the step aceuracy of 252Cf neutron afterloading radiotherapy machine and the dose deviation caused by step deviation.Methods EBT3 film was used to measure the steps of the 252Cf neutron source,and then the center of each 252Cf neutron source was identified by measuring the optical density value by using the ImageJ software.Double ion chambersmethod was used to measure the dose deviation dlue to the 252 Cf neutron source position shift.Results 252 Cf neutron source step accuracy may amount to 0.01 mm using EBT3 film measurement,when 252Cf neutron source position deviation is less than 3 mm,the dose deviation is less than 2.5%.Conclusions The study on the step accuracy and position deviation of the 252Cf neutron source can provide a reference for the quality control standard of the 252Cf neutron afterloading radiotherapy machine.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2017 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2017 Type: Article