Quantitative Analysis of Glucose Injection Content by Near Infrared Spectroscopy / 中国药师
China Pharmacist
;
(12): 1325-1328, 2017.
Article
in Chinese
| WPRIM
| ID: wpr-617583
ABSTRACT
Objective:
To quantitatively analyze the content of glucose in glucose injection by near infrared spectroscopy to control the quality of the product.Methods:
A quantitative model was established by near infrared reflectance spectroscopy and the injection came from pharmaceutical enterprises with different concentrations of glucose and the solution samples came from laboratories with different concentrations of glucose.The liquid sample accessories were selected, a quantitative model was established by a partial least squares method, and the effect of environmental temperature on the model was studied as well.Results:
According to the quantitative model, the correlation coefficient reached up to 0.999 3, and RMS deviation (RMSECV) was 0.077 4.In the verification test, the results of various liquid formulas containing glucose were similar to those of the laboratory results.The prediction error was less than 5% within the temperature range of 20-35℃.Conclusion:
The model of near-infrared partial least squares (PLS) can accurately predict the content of glucose in glucose products without sodium chloride, and can be used for the quality control of glucose intermediates and the finished products in the large infusion manufacturers
Full text:
Available
Index:
WPRIM (Western Pacific)
Type of study:
Prognostic study
Language:
Chinese
Journal:
China Pharmacist
Year:
2017
Type:
Article
Similar
MEDLINE
...
LILACS
LIS