Dialysis catheter fibrin sheath stripping: a useful technique after failed catheter exchange
Biomedical Imaging and Intervention Journal ; : 1-3, 2012.
Article in English | WPRIM | ID: wpr-625793
ABSTRACT
Fibrin sheath formation around long-term haemodialysis catheter is a common cause of failed dialysis access. Treatment options include pharmacological and mechanical methods. This paper reports a case of failed dialysis access due to fibrin sheath encasement. Pharmacologic thrombolysis, mechanical disruption using guide wire and catheter exchange had failed to address the issue. Eventually, fibrin sheath stripping using the loop snare technique was able to successfully restore the catheter function.

Full text: Available Index: WPRIM (Western Pacific) Language: English Journal: Biomedical Imaging and Intervention Journal Year: 2012 Type: Article