Your browser doesn't support javascript.
loading
Validity and Reliability of an Inertial Measurement Unit-Based 3D Angular Measurement of Shoulder Joint Motion
Journal of Korean Physical Therapy ; (6): 145-151, 2017.
Article in English | WPRIM | ID: wpr-646698
ABSTRACT

PURPOSE:

The purpose of this study was to investigate the validity and reliability of the measurement of shoulder joint motions using an inertial measurement unit (IMU).

METHODS:

For this study, 33 participants (32 females and 1 male) were recruited. The subjects were passively positioned with the shoulder placed at specific angles using a goniometer (shoulder flexion 0°–170°, abduction 0°–170°, external rotation 0°–90°, and internal rotation 0°–60° angles). Kinematic data on the shoulder joints were simultaneously obtained using IMU three-dimensional (3D) angular measurement (MyoMotion) and photographic measurement. Test-retest reliability and concurrent validity were examined.

RESULTS:

The MyoMotion system provided good to very good relative reliability with small standard error of measurement (SEM) and minimal detectable change (MDC) values from all three planes. It also presented acceptable validity, except for some of shoulder flexion, shoulder external rotation, and shoulder abduction. There was a trend for the shoulder joint measurements to be underestimated using the IMU 3D angular measurement system compared to the goniometer and photo methods in all planes.

CONCLUSION:

The IMU 3D angular measurement provided a reliable measurement and presented acceptable validity. However, it showed relatively low accuracy in some shoulder positions. Therefore, using the MyoMotion measurement system to assess shoulder joint angles would be recommended only with careful consideration and supervision in all situations.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Organization and Administration / Shoulder / Shoulder Joint / Photography / Reproducibility of Results / Range of Motion, Articular Limits: Female / Humans Language: English Journal: Journal of Korean Physical Therapy Year: 2017 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Organization and Administration / Shoulder / Shoulder Joint / Photography / Reproducibility of Results / Range of Motion, Articular Limits: Female / Humans Language: English Journal: Journal of Korean Physical Therapy Year: 2017 Type: Article