Your browser doesn't support javascript.
loading
Detection of SNR and geometric accuracy of MRI system based on ACR standards / 医疗卫生装备
Chinese Medical Equipment Journal ; (6): 74-77, 2017.
Article in Chinese | WPRIM | ID: wpr-662522
ABSTRACT
Objective To detect the SNR and geometric accuracy of MRI based on American College of Radiology (ACR)standards.Methods The SNR and geometric accuracy of Siemens Skyra 3.0T,Siemens Trio 3.0T and GE Excite HD 1.5T MRI were measured with ACR phantom,and the detection results were calculated according to the standards.Results The SNR values of Siemens Skyra 3.0T,Siemens Trio 3.0T and GE Excite HD 1.5T MRI were 589.98,438.50 and 277.12 respectively.Siemens Skyra 3.0T MRI had the values of geometric accuracy being-1.93% at X direction,-3.20% at Y direction and 0.68% at Z direction,Siemens Trio 3.0T MRI had the value of geometric accuracy being-0.87% at X direction,-2.33% at Y direction and 1.49% at Z direction,GE Excite HD 1.5T MRI had the values of geometric accuracy being 0.20% at X direction,-1.53% at Y direction and 1.69% at Z direction.Conclusion The detection of the SNR and geometric accuracy of MRI can effectively guarantee the image quality.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study Language: Chinese Journal: Chinese Medical Equipment Journal Year: 2017 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Diagnostic study Language: Chinese Journal: Chinese Medical Equipment Journal Year: 2017 Type: Article