ABSTRACT
Objective To explore application of our self-designed U-shaped transdermal surgical navigator based on three-dimensional reverse design in cannulated screwing for femoral neck fracture Methods A retrospective study was done of 5 patients with femoral neck fracture who had been treated from August 2016 to March 2017.They were 3 men and 2 women,aged from 33 to 60 years (average,46.5 years).Thin layer CT scanning was conducted on the patients' affected hip (at the left side in 2 cases and at the right side in 3).Based on their CT data,3D design software and 3D printing technique were used to manufacture the U-shaped transdermal surgical navigator and a real-sized model of the fracture.Operations were simulated using the U-shaped transdermal surgical navigator on the fracture model to determine the angle,depth and location of the screwing.Fixation of the fracture with cannulated screws was conducted after confirmation of satisfactory surgical parameters.The postoperative X-ray films of the patients were reviewed.Evaluations were conducted of the screw position,operating time,volume of operative bleeding,and number of intraoperative X-ray shots.Results The screw positions in the 5 patients were consistent with their preoperative design.Four patients had only one shot of fluoroscopy and one patient 3 shots.Their operating time ranged from 15 to 30 minutes,averaging 22.5 minutes;their bleeding ranged from 5 to 10 mL,averaging 7.5 mL.The 5 patients were followed up for an average of 4.5 months (from 3 to 6 months).By the Harris scoring at their final follow-ups,one was excellent,3 good and one fair.No complications like pain,functional impairment of the joint,or collapse or necrosis of the femoral head occurred during the follow-up.Conclusions Our self-designed U-shaped transdermal surgical navigator based on three-dimensional reverse design can significandy improve the accuracy and safety of the cannulated screwing for femoral neck fractures.The operation is minimally invasive,reducing operating time and intraoperative blood loss.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Observational study / Prognostic study Language: Chinese Journal: Chinese Journal of Orthopaedic Trauma Year: 2017 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Observational study / Prognostic study Language: Chinese Journal: Chinese Journal of Orthopaedic Trauma Year: 2017 Type: Article