Your browser doesn't support javascript.
loading
Correlation between measurement error of oxygen concentration and service time of ventilator's oxygen cell / 医疗卫生装备
Chinese Medical Equipment Journal ; (6): 83-85, 2017.
Article in Chinese | WPRIM | ID: wpr-668475
ABSTRACT
Objective To analyze the relationship between the measurement error of oxygen concentration and service time of oxygen cell of the ventilator. Methods The present situation of the ventilator was analyzed. Some cases of alarms for oxygen concentration failures were taken in statistical analysis on 3-a quality control detection results of the ventilators in some hospital. Results There was an approximately linear correlation between the measurement error of oxygen concentration and service time of oxygen cell, and the measurement error of oxygen concentration was increased significantly 430 days after the oxygen cell enabled. Conclusion The measurement error of oxygen concentration quality control could be used as a reference for performance evaluation and replacement of oxygen cell.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Medical Equipment Journal Year: 2017 Type: Article