ABSTRACT

Objective:

To establish a method for the determination of sodion in cefalotin sodium by ion chromatography and investi-gate the salt-forming rate of the products.

Methods:

A TSKgelSuper IC-CR cation exchange column (150 mm × 4. 6 mm, 3. 0 μm) was used. The mobile phase was the mixture of 2. 2 mmol·L-1 methanesulfonic acid and 1 mmol·L-1 18-crown-6-ether with the flow rate of 0. 8 ml·min-1 . The column temperature was 40℃ and the injection volume was 20μl. The detector was an electric conductiv-ity detector.

Results:

The linear correlation of sodion was good within the range of 3. 0-60. 0μg·ml-1(r=0. 999 9). The average re-covery was 99. 8%(RSD=0. 8%, n=9). The mole number ratio of sodion to cefalotin was within the range of 0. 97-1. 03. Conclu-sionThe method is specific, precise and accurate, and can be used in the determination of sodion in cefalotin sodium. The salt-form-ing rate of the 8 batches of samples is promising.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: China Pharmacist Year: 2015 Type: Article