Stress distribution of implanted complete overdentures retained with two types magnetic attachment / 实用口腔医学杂志
Journal of Practical Stomatology
; (6)2000.
Article
in Zh
| WPRIM
| ID: wpr-670726
Responsible library:
WPRO
ABSTRACT
Objective:To study the stress distribution of Magsoft (cushion type) and Magnedisk magnetic attachment. Methods: 2 overdentures were prepared on each of 5 edentulous jaw models. The overdentures were pared and retained with Magnedisk and Magsoft magnetic attachments respectively.Photoelastic analysis was used to study the stress distribution for the implanted complete overdentures. Results: Under the same load, around the implants, the stress under the overdentures retained with Magsoft was lower than that with Magnedisk. Compared with Magnedisk, the stress around the abutments under the overdenture retained with Magsoft were distributed in a more even manner. Conclusion: Magsoft magnetic attachment has better biomechanical characteristic than Magnedisk and can distribute stress more evenly throughout the whole supporting structure for its special cushioned design.
Full text:
1
Index:
WPRIM
Type of study:
Prognostic_studies
Language:
Zh
Journal:
Journal of Practical Stomatology
Year:
2000
Type:
Article