Comparison of conventional and supported liquid extraction methods for the determination of sitagliptin and simvastatin in rat plasma by LC-ESI-MS/MS / 药物分析学报
Journal of Pharmaceutical Analysis ; (6): 161-168, 2015.
Article in Chinese | WPRIM | ID: wpr-671991
ABSTRACT
Three extraction methods were compared for their efficiency to analyze sitagliptin and simvastatin in rat plasma by LC–MS/MS, including (1) liquid–liquid extraction (LLE), (2) solid phase extraction (SPE) and (3) supported liquid extraction (SLE). Comparison of recoveries of analytes with different extraction methods revealed that SLE was the best extraction method. The detection was facilitated with ion trap-mass spectrometer by multiple reactions monitoring (MRM) in a positive ion mode with ESI. The transitions monitored were m/z 441.1→325.2 for simvastatin, 408.2→235.1 for sitagliptin and 278.1→260.1 for the IS. The lower limit of quantification (LLOQ) was 0.2 ng/mL for sitagliptin and 0.1 ng/mL for simvastatin. The effective SLE offers enhanced chromatographic selectivity, thus facilitating the potential utility of the method for routine analysis of biological samples along with pharmacokinetic studies.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Journal of Pharmaceutical Analysis Year: 2015 Type: Article