Analytical quality-by-design approach for sample treatment of BSA-containing solutions / 药物分析学报
Journal of Pharmaceutical Analysis
;
(6): 27-32, 2015.
Article
in Chinese
| WPRIM
| ID: wpr-672143
ABSTRACT
The sample preparation of samples containing bovine serum albumin (BSA), e.g., as used in transdermal Franz diffusion cell (FDC) solutions, was evaluated using an analytical quality-by-design (QbD) approach. Traditional precipitation of BSA by adding an equal volume of organic solvent, often successfully used with conventional HPLC-PDA, was found insufficiently robust when novel fused-core HPLC and/or UPLC-MS methods were used. In this study, three factors (acetonitrile (%), formic acid (%) and boiling time (min)) were included in the experimental design to determine an optimal and more suitable sample treatment of BSA-containing FDC solutions. Using a QbD and Derringer desirability (D) approach, combining BSA loss, dilution factor and variability, we constructed an optimal working space with the edge of failure defined as Do0.9. The design space is modelled and is confirmed to have an ACN range of 8373%and FA content of 170.25%.
Full text:
Available
Index:
WPRIM (Western Pacific)
Type of study:
Prognostic study
Language:
Chinese
Journal:
Journal of Pharmaceutical Analysis
Year:
2015
Type:
Article
Similar
MEDLINE
...
LILACS
LIS