Validation of High Performance Liquid Chromatographic Method with UV Detector for the Determination of Di(2-ethylhexyl)Phthalate in Plasma in some Korean Male Workers / 대한산업의학회지
Korean Journal of Occupational and Environmental Medicine
;
: 70-78, 2005.
Article
in Korean
| WPRIM
| ID: wpr-67477
ABSTRACT
OBJECTIVES:
This study was conducted to validate a simple, rapid and sensitive reverse-phase high-performance liquid chromatographic method with UV detector (HPLC-UV) and present the plasma level of di(2-ethylhexyl)phthalate (DEHP) in some Korean male workers.METHODS:
HPLC-UV for quantification of plasma DEHP was validated by the following guideline from the Center for Drug Evaluation and Research (CDER)-calibration/standard curve, precision, accuracy and recovery. Plasma DEHP from 255 healthy Korean male workers aged from 30 to 60 years was analyzed by validated HPLC-UV method.RESULTS:
The calibration curve over the range 0~150 microgram/liter for the plasma DEHP standard solution showed linearity(r2=0.999). The limit of detection (LOD) and limit of quantification (LOQ) of plasma DEHP were 5.22 microgram/liter and 15.81 microgram/liter, respectively. The accuracy and precision for 2.5 microgram/liter of DEHP were acceptable in CDER guideline on the second and third day but not first day, and those for 50 microgram/liter and 150 microgram/liter of DEHP were acceptable on all three days(Ed-confirm this addition). The distribution of plasma DEHP level was skewed to the left and ranged from 0 to 18.9 microgram/liter. The plasma DEHP level was lower than 10 microgram/liter for 98 % of subjects and lower than 5 microgram/liter for 85 %. The geometric mean and standard deviation of plasma DEHP were 0.4 +/- 1.5 microgram/liter.CONCLUSIONS:
The HPLC-UV method for quantification of plasma DEHP was acceptable by CDER guideline. The plasma DEHP of 255 Korean male workers ranged from 0 to 18.9 microgram/liter and the distribution was skewed to the left.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Plasma
/
Calibration
/
Diethylhexyl Phthalate
/
Drug Evaluation
/
Limit of Detection
Limits:
Humans
/
Male
Language:
Korean
Journal:
Korean Journal of Occupational and Environmental Medicine
Year:
2005
Type: