Dual-sided reading versus single-sided reading:comparison of image quality and radiation dose between the two computed radiography systems / 中华放射学杂志
Chinese Journal of Radiology ; (12)2001.
Article in Chinese | WPRIM | ID: wpr-680034
ABSTRACT
Objective To assess and compare the difference in image quality and exposure dose between single-sided reading image plate(IP)and dual-sided reading IP.Methods A contrast-detail phantom CDRAD 2.0 was exposed by single-sided and dual-sided reading IP with different mAs sets.The entrance surface doses were recorded for all images.Images were then presented to two radiologists on a high resolution monitor of diagnosis workstation.The image quality figure(IQF)was measured for each image. Statistical analysis was performed using Spearman's correlation test and Wilcoxon signed-rank test to compare the difference in image quality and exposure dose between single-sided IP and dual-sided reading IP. Results With different tube current dosage of 5.6,12.0,20.0,25.0,and 40.0 mAs,IQF values of single-sided reading IP were 47.95,37.68,34.31,28.61,and 24.65,respectively,while those of dual- sided reading IP were 38.83,29.81,29.65,25.16,and 21.43,respectively.The IQF difference between them showed statistical significance(P