Calibration and Radiation Survey of High Dose Rate Remote Afterloading System / 대한치료방사선과학회지
Journal of the Korean Society for Therapeutic Radiology
;
: 101-112, 1995.
Article
in Korean
| WPRIM
| ID: wpr-6863
ABSTRACT
High Dose Rate Remote Afterloading system was installed at Wonkwang Universi-ty Hospital in January 1994. In this report, the calibration of a Gammamed 12-i High Dose Rate Remote Afterloading system and the radiation survey around the facility after design and construct a shieding room are discussed. The radiation survey of the facility indicates that the use of ordinary concrete shielding of existing room will provide adequate shielding. Also, the methodologies for performing source calibra-tion are presented.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Calibration
/
Carboxymethylcellulose Sodium
Type of study:
Prognostic study
Language:
Korean
Journal:
Journal of the Korean Society for Therapeutic Radiology
Year:
1995
Type:
Article
Similar
MEDLINE
...
LILACS
LIS