Annual Report on External Quality Assessment in Urinalysis in Korea (2004) / 임상검사와정도관리
Journal of Laboratory Medicine and Quality Assurance
; : 59-83, 2005.
Article
in Ko
| WPRIM
| ID: wpr-68695
Responsible library:
WPRO
ABSTRACT
Three external quality assesment trials which composed of 16 control materials(12 chemical materials and four microscopic photographs of urinary sediment) for interlaboratory quality control assesment in urinalysis were performed with 446, participants, in each, in the year of 2004. The response rate were 93.4% (422/452), 91.7% (411/448) and 91.7% (410/447), in the first , the second and the third trials, in each. The test items include pH, glucose, protein, ketone, bilirubin, blood, urobilinogen, nitrite, leukocyte estrase, specific gravity and four microscopic photographs of urinary sediment. The survey results are summarized as follows: 1.The chemical quality control test in urinalysis revealed generally good concordance. 2.The percentage of using urinalysis analyzer was slightly increased as 87.9% and the distribution of using reagent strip was similar to the previous year. 3.The percentage of response rate of microscopic photographs of urinary sediment was 80.7% and the percentage of good performance of these tests ware 85.8% to 95.8%.
Key words
Full text:
1
Index:
WPRIM
Main subject:
Quality Control
/
Reagent Strips
/
Specific Gravity
/
Urobilinogen
/
Bilirubin
/
Urinalysis
/
Equidae
/
Glucose
/
Hydrogen-Ion Concentration
/
Korea
Country/Region as subject:
Asia
Language:
Ko
Journal:
Journal of Laboratory Medicine and Quality Assurance
Year:
2005
Type:
Article