ABSTRACT
Spontaneous fracture of the Silastic shunt catheter is a rare complication of lumboperitoneal shunt. A review of the literature revealed only one case in which spontaneous fracture occurred after this procedure. The authors report a case in which fracture of a lumboperitoneal shunt catheter occurred within the interspinous ligament probably due to repeated tension caused by flexion and extension during lumbar motion. The fractured proximal catheter was located entirely within the intrathecal space exclusively from the level of L1 to S1, and caused incomplete cauda equina syndrome.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Polyradiculopathy / Catheters / Fractures, Spontaneous / Ligaments Language: Korean Journal: Journal of Korean Neurosurgical Society Year: 1998 Type: Article