Your browser doesn't support javascript.
loading
Overall Evaluation and Management of 8 Instruments in Emergency Laboratory Using 6σ Quality Standard / 现代检验医学杂志
Journal of Modern Laboratory Medicine ; (4): 146-150, 2018.
Article in Chinese | WPRIM | ID: wpr-696231
ABSTRACT
Objective To evaluate and improve overall quality detection of 8 instruments in emergency laboratory using 6σ quality standard.Methods In 2017,collected both internal quality control (IQC) and external quality assessment (EQA) data of 46 assay items from 8 instruments in their laboratory,and then calculated their σ values and total allowed errors ac cording to CLIA'88 standards.By comparison of their quality goal index (QGI) between first and second half year of 2017,successfully elevated assayquality using strategy (man,machine,material,method,measurment and enviroment).Results For all 46 assay items (except NT-proBNP without second EQA data),made better improvement of detection of 34 (75.6 % of 6σ qualified),38(84.4% of 5σ qualified),40(88.9% of 4σ qualified),44 (97.8% of 3σ qualified) and 1 (2.2% of 2~3σ qualified) items,respectively whereas there were 27 (58.7% of 6σ qualified),30 (65.2% of 5σ qualified),35 (76.1% of 4σ qualified),41 (89.1% of 3σ qualified) and 5 (10.9% of 2~3σ qualified) items before improvement.Conclusion Using 6σ quality standard,could make more progress in the laboratory for many respectsfast and accurate clinical reports,reducing reinspection rate,and increasing patient's satisfactory,etc.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Practice guideline Language: Chinese Journal: Journal of Modern Laboratory Medicine Year: 2018 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Practice guideline Language: Chinese Journal: Journal of Modern Laboratory Medicine Year: 2018 Type: Article