ABSTRACT
Objective To measure absorbed dose and verify two-dimensional dose distribution from IMRT MLC using thermoluminescence dosimeters (TLDs) and films.Methods The teflon phantom was scanned by CT and its images were transmitted to TPS for planning.The 6 Gy-corresponding MUs were calculated at 90 cm SSD and 10 cm depth and on 5 cm × 5 cm radiation field.All the plans were implemented at 7 medical linear accelerators,which were produced by Varian,Elekta and Siemens and selected from 7 third-grade first-class hospitals according to the technical conditions of various regions in Sichuan province.The homogeneous solid phantoms used in hospitals were covered by 30 cm × 30 cm and 25 cm × 25 cm films.Then,the films were covered by thicker-than-20 cm phantoms.Finally,the plans were implemented by aligning the center of beams to the films center.Results The relative deviations of the measured absorbed dose to TPS-planned dose were 1.4%,3.7%,-2.5%,-0.3%,4.9%,4.9%,5.0% for TLDs and 4.7%,4.3%,1.5%,3.9%,-1.6%,3.3%,-1.3% for films,respectively,all consistent with the limit of less than 5%.The passing rates of 2D dose distribution (3 mm/3%) were 99.9%,98.5%,98.5%,97.9% and 70.0% for 5 accelerators,with only one not consistent with the requirements.Conclusions It is convenient to measure the absorbed dose to photon beam field and verify two-dimensional dose distribution using TLDs and films,which can provide quality assurance for radiation treatment plans.

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2018 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2018 Type: Article