Your browser doesn't support javascript.
loading
Influence on image quality in low-dose colonography with adaptive statistical iterative reconstruction algorithm / 中华放射医学与防护杂志
Chinese Journal of Radiological Medicine and Protection ; (12): 311-316, 2018.
Article in Chinese | WPRIM | ID: wpr-708061
ABSTRACT
Objective To evaluate the feasibility of adaptive statistical iterative reconstruction-V (ASIR-V) to improve image quality in low-dose CT colonography.Methods A series of thirty artificial polyps were established by ligation in an isolated segment of porcine colon.Volume data was acquired on Revolution CT scanner (GE,USA) with High Definition scan mode and different scan parameter combinations120 kVp with different mAs (10,30,50,70,90,100,120,140,160,180,200,220,240,260,respectively).Images were reconstructed with six different ASIR-V levels of 0 (filtered back projection,FBP),10%,30%,50%,70% and 90%.Two radiologists were blinded to measure and analyze the objective data independently,including image noise (SD),signal-to-noise ratio (SNR) and contrast-to-noise ratio (CNR).The kappa test was used to assess interobserver agreement in subjective image quality score.ICC test was used to examine the consistency of the measurements between two observers.SD,SNR,CNR were performed for statistical analysis in different tube current and different levels of ASIR-V using variance analysis (ANOVA).Results Interobserver agreement for subjective image quality score was good with a kappa value of 0.683.The variation of the tube current(r =0.734,P =0.000) and ASIR-V level(r =0.220,P =0.044) is related to the subjective score of image quality.Under the condition of the same tube current,image quality score of 50% ASIR-V reconstruction was the highest.Two objective data consistency is good.The differences of image noise (F =423.58,P < 0.05),SNRs(F =124.26,P < 0.05) and CNRs (F =1 030.17,P < 0.05) of different tube current and different levels of ASIR-V reconstruction were statistically significant.In the same tube current,with increased levels of ASIR-V,image noise reduced,CNRs increased.Only in 10,120,140,160,220,240,260 mA,the differences of SNRs were statistically significant(F =8.75-31.36,P < 0.05).For the same level of ASIR-V reconstruction,with the increase of tube current,the image noise decreased,SNR and CNR increased gradually.Conclusions In the CT colonography,the application of ASIR-V algorithm can significantly reduce the noise and enhance the image contrast noise ratio and improve image quality.ASIR-V algorithm with 50% has better performance in reducing CT image noise.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Prognostic study Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2018 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Type of study: Prognostic study Language: Chinese Journal: Chinese Journal of Radiological Medicine and Protection Year: 2018 Type: Article