Planning target volume margin based on the image-guided radiotherapy in nasopharyngeal carcinoma / 肿瘤研究与临床
Cancer Research and Clinic
;
(6): 180-183,189, 2018.
Article
in Chinese
| WPRIM
| ID: wpr-712791
ABSTRACT
Objective To analyze set-up errors for irradiation of nasopharyngeal carcinoma by using kilo-voltage cone beam computed tomography (kV-KBCT) scanning, and to calculate the external margin from planning target volume (PTV) of nasopharyngeal carcinoma. Methods A total of 150 nasopharyngeal carcinoma patients in Meizhou People's Hospital from December 2014 to November 2016 were treated by image-guided radiation therapy (IGRT), kV-KBCT, CT image scanning matcthed by bone and grey alignment. PTV of nasopharyngeal carcinoma was also calculated. Results According to PTV formula, the external distance before radiotherapy guided by grey alignment was 0.5 mm in X-axis, 0.4 mm in Y-axis, 0.8 mm in Z-axis. While the distance was 0.1 mm in X-axis, 0.5 mm in Y-axis, 1.7 mm in Z-axis guided by bone alignment. After radiotherapy, the external distance guided by grey alignment was 0.4 mm in X-axis, 0.5 mm in Y-axis, 0.9 mm in Z-axis. While the distance was 0.1 mm in X-axis, 0.9 mm in Y-axis, 2.0 mm in Z-axis guided by bone alignment. There was no significant difference in set-up errors of 3 directions and 2 aligned ways before and after treatment.Conclusions The PTV within 3 mm is safe when IGRT is used for directing radiotherapy of nasopharyngeal carcinoma,and kV-KBCT is an effective image equipment.
Full text:
Available
Index:
WPRIM (Western Pacific)
Language:
Chinese
Journal:
Cancer Research and Clinic
Year:
2018
Type:
Article
Similar
MEDLINE
...
LILACS
LIS