Validation of the mobile wireless digital automatic blood pressure monitor using the cuff pressure oscillometric method, for clinical use and self-management, according to international protocols
Biomedical Engineering Letters
;
(4): 399-404, 2018.
Article
in English
| WPRIM
| ID: wpr-717983
ABSTRACT
The purpose of this study was to evaluate the accuracy of a mobile wireless digital automatic blood pressure monitor for clinical use and mobile health (mHealth). In this study, a manual sphygmomanometer and a digital blood pressure monitor were tested in 100 participants in a repetitive and sequential manner to measure blood pressure. The guidelines for measurement used the Korea Food & Drug Administration protocol, which reflects international standards, such as the American National Standard Institution/Association for the Advancement of Medical Instrumentation SP 10 1992 and the British Hypertension Society protocol. Measurements were generally consistent across observers according to the measured mean ± SD, which ranged in 0.1 ± 2.6 mmHg for systolic blood pressure (SBP) and 0.5 ± 2.2 mmHg for diastolic blood pressure (DBP). For the device and the observer, the difference in average blood pressure (mean ± SD) was 2.3 ± 4.7 mmHg for SBP and 2.0 ± 4.2 mmHg for DBP. The SBP and DBP measured in this study showed accurate measurements that satisfied all criteria, including an average difference that did not exceed 5 mmHg and a standard deviation that did not exceed 8 mmHg. The mobile wireless digital blood pressure monitor has the potential for clinical use and managing one's own health.
Full text:
Available
Index:
WPRIM (Western Pacific)
Main subject:
Self Care
/
Blood Pressure
/
Blood Pressure Monitors
/
Telemedicine
/
Sphygmomanometers
/
Hypertension
/
Korea
/
Methods
Type of study:
Practice guideline
Country/Region as subject:
Asia
Language:
English
Journal:
Biomedical Engineering Letters
Year:
2018
Type:
Article
Similar
MEDLINE
...
LILACS
LIS