Your browser doesn't support javascript.
loading
Development of QA Phantom Prototype for Imaged Based Radiation Treatment System / 의학물리
Korean Journal of Medical Physics ; : 120-124, 2008.
Article in Korean | WPRIM | ID: wpr-7198
ABSTRACT
In this study, we developed the protopype of QA phantom for image QA including an additional component for image based radiation treatment system. The new phantom considered two main parts Image quality and fusion accuracy. Image quality part included for daily CT number linearity and spatial resolution, and fusion accuracy part designed to simulate a simple translation-rotation setting. The CT scans of the phantom obtained from conventional CT, MVCT of Tomotherapy unit, and both image sets were satisfied the recommendation of spatial resolution. This phantom was simple and efficient for daily imaging QA, and it is important to provide a new concept of verification of image registration.

Full text: Available Index: WPRIM (Western Pacific) Type of study: Practice guideline Language: Korean Journal: Korean Journal of Medical Physics Year: 2008 Type: Article