Your browser doesn't support javascript.
loading
Patient Radiation Exposure During Diagnostic and Therapeutic Procedures for Intracranial Aneurysms: A Multicenter Study / 신경중재치료의학
Neurointervention ; : 78-85, 2016.
Article in English | WPRIM | ID: wpr-730322
ABSTRACT

PURPOSE:

To assess patient radiation doses during cerebral angiography and embolization of intracranial aneurysms across multi-centers and propose a diagnostic reference level (DRL). MATERIALS AND

METHODS:

We studied a sample of 490 diagnostic and 371 therapeutic procedures for intracranial aneurysms, which were performed at 23 hospitals in Korea in 2015. Parameters including dose-area product (DAP), cumulative air kerma (CAK), fluoroscopic time and total angiographic image frames were obtained and analyzed.

RESULTS:

Total mean DAP, CAK, fluoroscopy time, and total angiographic image frames were 106.2 ± 66.4 Gy-cm2, 697.1 ± 473.7 mGy, 9.7 ± 6.5 minutes, 241.5 ± 116.6 frames for diagnostic procedures, 218.8 ± 164.3 Gy-cm², 3365.7 ± 2205.8 mGy, 51.5 ± 31.1 minutes, 443.5 ± 270.7 frames for therapeutic procedures, respectively. For diagnostic procedure, the third quartiles for DRLs were 144.2 Gy-cm² for DAP, 921.1 mGy for CAK, 12.2 minutes for fluoroscopy times and 286.5 for number of image frames, respectively. For therapeutic procedures, the third quartiles for DRLs were 271.0 Gy-cm² for DAP, 4471.3 mGy for CAK, 64.7 minutes for fluoroscopy times and 567.3 for number of image frames, respectively. On average, rotational angiography was used 1.5 ± 0.7 times/session (range, 0-4; n=490) for diagnostic procedures and 1.6 ± 1.2 times/session (range, 0-4; n=368) for therapeutic procedures, respectively.

CONCLUSION:

Radiation dose as measured by DAP, fluoroscopy time and image frames were lower in our patients compared to another study regarding cerebral angiography, and DAP was lower with fewer angiographic image frames for therapeutic procedures. Proposed DRLs can be used for quality assurance and patient safety in diagnostic and therapeutic procedures.
Subject(s)

Full text: Available Index: WPRIM (Western Pacific) Main subject: Cerebral Angiography / Angiography / Fluoroscopy / Intracranial Aneurysm / Radiation Exposure / Patient Safety / Korea Type of study: Controlled clinical trial / Diagnostic study Limits: Humans Country/Region as subject: Asia Language: English Journal: Neurointervention Year: 2016 Type: Article

Similar

MEDLINE

...
LILACS

LIS

Full text: Available Index: WPRIM (Western Pacific) Main subject: Cerebral Angiography / Angiography / Fluoroscopy / Intracranial Aneurysm / Radiation Exposure / Patient Safety / Korea Type of study: Controlled clinical trial / Diagnostic study Limits: Humans Country/Region as subject: Asia Language: English Journal: Neurointervention Year: 2016 Type: Article