Design of a portable SSVEP signal acquisition system / 国际生物医学工程杂志
Article in Zh | WPRIM | ID: wpr-751617
Responsible library: WPRO
ABSTRACT
Objective To design a portable electroencephalography(EEG) acquisition system to acquire and analysis steady-state visual potentials (SSVEP). Methods The microprocessor MSP432P401 series MCU was used to control the high-performance integrated analog front end ADS1299 to realize the acquisition, amplification and analog-to-digital (AD) conversion of EEG signals. The digital EEG signal is sent to the host computer for processing by WIFI. Spontaneous EEG signals and steady-state visually evoked EEG signals from 3 healthy subjects were collected to verify system performance. Results The collected signal had a clear α-wave rhythm of closed-eye spontaneous EEG signals. The power spectrum density shows that the steady-state visually induced EEG signal frequency and harmonic frequency peak at the corresponding stimulation frequency, indicating that the system works normally and the performance is good. Conclusions The designed portable EEG acquisition system can accurately collect the spontaneous and induced EEG signals of the human body, which provides technical support for the clinical application of SSVEP technology.
Key words
Full text: 1 Index: WPRIM Language: Zh Journal: International Journal of Biomedical Engineering Year: 2019 Type: Article